Frequently Asked Questions B1520A - Capacitance range and resolution for CV measurement. Capacitance range and resolution for CV measurement. MFCMU, B1520A, does not measure a capacitance direct. Therefore, we do not have the definition of Capacitance resolutions and ranges. MFCMU measures an impedance and then calculate to capacitance. How to calculate impedance to the capacitance, please see the user guide. http://literature.cdn.keysight.com/litweb/pdf/B1500-90000.pdf#page=73 P.73 Measurement range: The impedance ranges are here in the user guide too. http://literature.cdn.keysight.com/litweb/pdf/B1500-90000.pdf#page=75 P.75 Table 2-16 Measurement Range for Fixed Ranging Mode