Can I change the *IDN? response of my X Series Signal Analyzer for code compatibility?
Can I change the *IDN? response of my X Series Signal Analyzer for code compatibility?
Frequently Asked Questions (FAQs)SummaryThis article shows how to change the *IDN? response of the N9000A/10A/20A/30A/00B/10B/20B/30B/40B/41B X series of analyzers. This is sometimes useful for programs that were written for other units, and are hard coded to expect another instrument.QuestionCan I change the *IDN? response of my X Series Signal Analyzer for code compatibility?AnswerYes you can. There are 3 choices: Setting Format of *IDN? command Factory Keysight Technologies, model, serial, software versionAgilentAgilent Technologies, model, serial, software versionUserAnything you want! This is editable. On your X Series Signal Analyzer, press the System Hardkey. Then: On the multitouch units: You should see: Some examples of *IDN? query with the different setting: Factory: Keysight Technologies,N9020B,MY55470107,A.24.57 Agilent: Agilent Technologies,N9020B,MY55470107,A.24.57 User: Hewlett Packard, E4407B, MY55470107,A.24.57 On the A models (legacy Gui) press System, I/O Config, System IDN Response to find the equivalent settings.