Can I Perform 1153A Probe Calibration With Its 100:1 Attenuator?
Frequently Asked Questions (FAQs)SummaryNo.QuestionCan I perform 1153A probe calibration with its 100:1 attenuator?AnswerWe cannot perform 1153A probe calibration with its 100:1 attenuator. This is because the oscilloscope's calibration output is unable to provide a signal of sufficient dynamic range once the 100:1 attenuator is attached to 1153A. However, we can still calibrate 1153A with the 10:1 attenuator. Subsequently, we can change to the 100:1 attenuator for high voltage measurement.