Can I Restore BenchLink DataLogger (BLDL1) Datasets?

Can I Restore BenchLink DataLogger (BLDL1) Datasets?

Yes. Follow the steps below to retrieve your datasets generated by BLDL1 (make sure to copy the Y*.* and X*.* files somewhere else for backup):Copy the Y*.* and X*.* into C:\bench\dlogger\data32Right click on those two files and view the PropertiesUncheck the Read-only then press OKOpen Keysight BenchLink Data Logger 1.5File -> Modify/Delete...Delete that data if it already exists; otherwise, press Cancel and Exit BenchLinkOpen C:\bench\dlogger\util32\Dldbx32.exeRetrieve the data then turn off Dldbx32.exeOpen BenchLink againFile -> Open -> open one of the Data from the Setup dropdown (+ sign)If the above steps do not work, try the following:Close all BLDL1 applicationsRun C:\bench\dlogger\util32\dlutil32.exe programPress the Clean button and exit when it's done cleaningOpen the C:\bench\dlogger\data32 folder and make sure the X and Y files are in thereRun C:\bench\dlogger\util32\Dldbx32.exe program and press RestoreClose the Dldbx32.exe program and open BLDL1 and open that configuration with the datasetAt the Data Browser window, press File/Export Data... and export the dataset from thereIf you have tried both ways and it still doesn’t retrieve the data, there is no other way to restore because the data file is corrupted. We recommend using BLDL3 instead.

Was this answer helpful?