Can I Use TRM Calibration Method On E5071C?
Can I Use TRM Calibration Method On E5071C?
Frequently Asked Questions (FAQs)SummaryYes, E5071C could support TRM calibration kit. However, you need to add the TRM calibration kit to E5071C as user calibration kit.QuestionCan I Use TRM Calibration Method On E5071C?AnswerYes, E5071C could support TRM calibration kit. However, you need to add the TRM calibration kit to E5071C as user calibration kit. If you are not sure the format of TRM calibation kit, please use "predefined 85052C calibration kit" and then modify it. 85052C is a 3.5mm TRL kit. Below 2 GHz, the kit uses load standards instead of lines. So, TRM is implemented in this frequency range" See AlsoThe FAQ is published at https://community.keysight.com/thread/4359 Please refer to E5071C on line help for Modifying Calibration Kit Definition. http://ena.support.keysight.com/e5071c/manuals/webhelp/eng/measurement/calibration/advanced_calibrations/changing_the_calibration_kit_definition.htm