keysight_articles upload_to_live

Can I convert the AC line voltage from 230 to 115 VAC for my E3631xA DC power supply?

Can I convert the AC line voltage from 230 to 115 VAC for my E3631xA DC power supply?

Frequently Asked Questions (FAQs)SummaryYes, this FAQ discusses AC line voltage conversion for the E3631xA series DC power supplies.QuestionCan I convert the AC line voltage from 230 to 115 VAC for my E3631xA DC power supply?  AnswerYes, changing the AC Line Voltage selection for an E36300 Series DC power supply is a very quick and easy task.  You may select 115 VAC, 230 VAC, or 100 VAC nominal line voltage. All settings are valid for +/- 10% of the nominal line voltage and frequency (47 - 63 Hz) To change the input AC voltage selector on the power supply, use the two AC selector switches on the underside of the power supply as shown. For example, to select 230 V, slide the left switch to the left and the right switch to the right, as illustrated in the diagram below the switches.  To select 115 V, slide both switches to the left, and to select 100 V, slide the two switches towards each other. After changing the line voltage selection switches, be sure to check and use the correct fuse. The following table provides the fuse rating, description, and Keysight part number that you should use with each of the power supply model and option. Option 0E9 is for nominal 100 VAC, 0EM is for 115 VAC, and 0E3 is for 230 VAC: To install the correct fuse, follow the three steps shown below:    AttachmentsSee AlsoE36300 Series Programmable DC Power Supplies - User’s Guide, p.23-24, Edition 5, 25-Nov-2019  https://literature.cdn.keysight.com/litweb/pdf/E36311-90001.pdf

Was this answer helpful?

0 found this article helpful