33120A: Clearing Memory

33120A: Clearing Memory

Calibration Constants:The calibration constants cannot be cleared except during a calibration.Calibration String:The calibration string allows you to store customized calibration information about your function generator. To clear the calibration string, simply overwrite the old string by entering new information (or null string) using the CALibration:STRing command. For more information, refer to chapter 4 in the Keysight 33120A User's Guide.Instrument States:You can store up to four different instrument states (numbered 0,1,2, and 3) in non-volatile memory. To clear a previously stored instrument state, use the MEMory:STATe:DELete{0|1|2|3} command. For more information, refer to chapter 4 in the Keysight 33120A User's Guide.User-Defined Arbitrary Waveform Memory:You can delete the waveform currently in volatile memory or any of the four user-defined waveforms in non-volatile memory. Note that you cannot delete the arbitrary waveform that is currently being output nor can you delete any of the five built-in arbitrary waveforms. Use the DATA:DELete command to delete selected arbitrary waveforms or the DATA:DELete:ALL command to delete all user-defined arbitrary wavforms. For more information, refer to chapter 4 in the Keysight 33120A User's Guide.Revision 1.0, 17 Dec 98 - dah