Can I increase measurement speed when using the N897xA NFA Series remotely?
Can I increase measurement speed when using the N897xA NFA Series remotely?
Switch the NFA display off using DISP:ENAB OFF if this is applicable.Set the NFA to Single sweep using INIT:CONT OFF and use the 'Read' commands to retrieve data.e.g. READ:SCAL:CORR:GAIN? will initiate a fixed frequency gain measurement and retrurn the result.If more than one parameter is being measured use the 'Read' command to return the first parameter then the 'Fetch' command to return the remaining parameters.