33120A: Replacing the Output Protection Fuse

33120A: Replacing the Output Protection Fuse

The output protection fuse is located inside the function generator. This fuse is a thru-hole soldered 500 mA part (Keysight part number 2110-0716C). The fuse is located near the output connector (J801) on the main PC board. You will need to disassemble the function generator to replace this fuse (use a TORX T-15 driver to remove the screws located on the rear panel).This part may be obtained from Keysight Parts Ordering at 877-447-7278.Revision 1.0 13 DEC 99 jtm